Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa HR1300: CAL ADJ /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 117 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # yo13_52.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 1-hour - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 HEAD {CONNECTIONS} 1.006 PIC yo13_52 1.007 OPBR Are Channels 1, 3 & 5 connected properly? 1.008 JMPF 1.006 1.009 DISP Save settings to IC card or print "SET UP LIST" 1.010 DISP Turn the UUT power off. 1.010 DISP 1.010 DISP Set switch 1,2,4,8 of the DIP switch on the rear panel 1.010 DISP to ON. 1.011 DISP Hold LOWER DISPLAY key depressed and turn on the UUT. 1.011 DISP 1.011 DISP Hold LOWER DISPLAY key until 'SET UP MODE=RECORD' 1.011 DISP appears on the display. 1.012 DISP Select 'A/D' using function key and press ENTRY key. 1.012 DISP 'A/D MODE=DC DISP' appears on the display. 1.012 DISP 1.012 DISP Select 'DC AVE' using function key and press ENTRY. 1.012 DISP '20mV ZERO' appears on the display. 1.013 DISP Press ENTRY. The UUT adjusts 0mV automatically and 1.013 DISP then 'A/D MODE=DC AVERAGE' appears on the display. 1.014 DISP Press ENTRY. 1.014 DISP 'AVE=20mV ZERO' appears on the display. 1.015 DISP Select '0ohm:1-Z' using function key and press ENTRY. 1.015 DISP 1.015 DISP The UUT adjusts 0 Ohms automatically and then 1.015 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.016 DISP Press ENTRY. 1.016 DISP 'AVE=0ohm:1mA' appears on the display. 1.017 ASK- U # Insure lead resistance is less than 10 Ohms. 1.018 5520 100.0Z S 2W 1.019 ASK+ U 1.020 DISP Select '100ohm:1' using function key and press ENTRY. 1.020 DISP 1.020 DISP The UUT adjusts 100ohm automatically and then 1.020 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.021 DISP Press ENTRY. 1.021 DISP 'AVE=100ohm:1mA' appears on the display 1.022 DISP Select '100:1H' using function key and press ENTRY. 1.022 DISP 1.022 DISP The UUT adjusts 100 Ohm automatically and then 1.022 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.023 DISP Press ENTRY. 1.023 DISP 'AVE=100ohm:1mA-H' appears on the display. 1.024 DISP Select '0ohm:2-Z' using function key and press ENTRY. 1.024 DISP 1.024 DISP The UUT adjusts 0 Ohm automatically and then 1.024 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.025 DISP Press ENTRY. 1.025 DISP 'AVE=0ohm:2mA' appears on the display 1.026 ASK- U # Insure lead resistance is less than 10 Ohms. 1.027 5520 100.0Z S 2W 1.028 ASK+ U 1.029 DISP Select '100ohm:2' using function key and press ENTRY. 1.029 DISP 1.029 DISP The UUT adjusts 100 Ohm automatically and then 1.029 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.030 DISP Press ENTRY. 1.030 DISP 'AVE=100ohm:2mA' appears on the display 1.031 DISP Select '100:2H' using function key and press ENTRY. 1.031 DISP 1.031 DISP The UUT adjusts 100 Ohm automatically and then 1.031 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.032 DISP Select 'WRITE' using function key and press ENTRY. 1.032 DISP 1.032 DISP 'WRITE OK=NO' appears on the display. 1.032 DISP 1.032 DISP Select 'YES' using function key and press ENTRY. 1.032 DISP 'A/D MODE=WRITE' appears on the display. 1.033 DISP Select 'ESC' using function key and press ENTRY. 1.033 DISP '***A/D SET***' appears on the display. 1.034 DISP Press ENTRY. 1.034 DISP 'SET UP MODE=A/D-ADJ' appears on the display. 1.034 DISP 1.034 DISP Select 'END' using function key and press ENTRY. 1.034 DISP 1.034 DISP The UUT returns to normal operation. 1.035 END